Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Cut microbore tubing into 48'' lengths, one per port. If a dial-a-wave device is used, 96" lines are necessary to compensate for the increased height of the linear actuators
  2. Remove the plunger from a 10ml syringe
    • Note: For dial-a-wave devices, 25ml or 60ml syringes are used for the media reservoirs. The protocol is the same, but the larger syringes are required to reduce the impact of media depletion.
  3. Attach a luer stub to the syringe body
  4. Insert the luer stub into one end of the microbore tubing
  5. Insert a straight pin into the other end of the microbore tub
  6. Decant 10ml of sterile water into the syringe, letting it run down the side of the syringe body
  7. Flick the luer stub to clear any bubbles
  8. Let the water flow through the tube until all of the bubbles have been cleared
  9. Top the syringe with a square of parafilm
  10. Use the pin to poke a hole in the parafilm, and store the tube until it is time to load the chip
  11. Repeat steps 2 to 10 for remaining water syringes and media syringe(s)
  12. Once the cells have reached OD600 of ~0.1, transfer 10ml of cells to a 15ml conical tube. Spin at ~1500xg (3500RPM in a Sorvall SS-34 rotor) for 5 minutes.
  13. While the cells are spinning down, turn on the fluorescence source for the microscope so that the bulb can reach operating temperature.
    • Note: During the first 5-10 minutes of on-time, the intensity of the fluorescence source will fluctuate until the bulb reaches operating temperature. Therefore, it is important to allow a sufficiently long warm up time. Once the source is on, leave the output at 0\% 0% to prevent accidental exposure when configuring the run.
  14. Decant and discard the supernatant. Add 10ml fresh LB media with antibiotics and resuspend the cells by vortexing briefly
  15. Repeat steps 2 to 10 with the cell syringe

...