Original Hg light source was replaced with LED light source. Equivalent powers were determined using Chroma fluorescent slides at 60x using the test exposure times.

Fluorescence

Channel

Test Exposure

Time

Average Camera Readout

at 10% of Hg Lamp

Percent Power of

LED Source

CFP300 ms~1,4308%
GFP1 ms~4,0004%
YFP10 ms~4,13012%
mCherry1 ms~10,900-11,00019%